DESCRIPTION
1-5
1.3
FUNCTIONAL DESCRIPTION
A functional block diagram of a typical GC System installation is shown in Figure 1-1. A sample
of the gas to be analyzed is taken from the process stream by a sample probe installed in the process
line. The sample passes through a sample line to the sample conditioning system where it is filtered
or otherwise conditioned. After conditioning, the sample flows to the Analyzer for separation and
detection of the components of the gas.
The chromatographic separation of the sample gas into its components is accomplished in the
Analyzer in the following manner. A precise volume of sample gas is injected into one of the unit's
analytical columns. The column contains a stationary phase (packing) that is either an active solid
(adsorption partitioning) or an inert solid support that is coated with a liquid phase (absorption
partitioning). The gas sample is moved through the column by means of a mobile phase (carrier
gas). Selective retardation of the components of the sample takes place in the column that causes
each component to move through the column at a different rate. This action separates the sample
into its constituent components.
A detector located at the outlet of the analytical column senses the elution of components from the
column and produces electrical outputs proportional to the concentration of each component.
Outputs from the Analyzer detectors are amplified in the Analyzer electronics, then transmitted to
the GC Controller for further processing. (See also, Section 1.6, "Theory of Operation.")
Output from the GC Controller is normally displayed on a remotely located personal computer (PC)
or a printer. Connection between the GC Controller and the PC can be accomplished via a direct
serial line or via the Modbus-compatible communication interface.
Multiple chromatograms may be displayed on the PC monitor, and compared or contrasted with
separate color schemes. This allows a stored chromatogram to be compared/contrasted with a
current or another stored chromatogram. This could be of great assistance when changing
parameters or isolating a problem.
Use of a PC for configuration and troubleshooting procedures is essential in most instances. (Basic
operations can also be performed from a keyboard and liquid crystal display that are built into certain
versions of the GC Controller.) The PC may be remotely connected via telephone, radio or satellite
communications. Once installed and configured, the GC System can operate independently for long
periods of time.
Summary of Contents for Daniel Danalyzer 1000
Page 2: ......
Page 18: ...SEP 2005 MODEL 1000 DESCRIPTION 1 6 Figure 1 1 GC System Functional Block Diagram ...
Page 38: ...SEP 2005 MODEL 1000 EQUIPMENT DESCRIPTION 2 6 Figure 2 1 Analyzer Components ...
Page 146: ...SEP 2005 MODEL 1000 INSTALLATION AND SETUP 3 92 Figure 3 37 Analyzer Components ...
Page 188: ...SEP 2005 MODEL 1000 MAINTENANCE 5 20 Figure 5 2 Analyzer Components ...
Page 222: ...SEP 2005 MODEL 1000 MAINTENANCE 5 54 This page intentionally left blank ...
Page 228: ...SEP 2005 MODEL 1000 LIST OF APPENDICES A ii This page intentionally left blank ...
Page 260: ...SEP 2005 MODEL 1000 SETTING SOLENOID PURGE FLOWS E 4 This page intentionally left blank ...
Page 282: ...SEP 2005 MODEL 1000 ANALYZER DRAWINGS Addendum 1 2 This page intentionally left blank ...
Page 283: ......
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......
Page 311: ......
Page 312: ......
Page 313: ......
Page 314: ......
Page 316: ......
Page 318: ......
Page 319: ......