SEP 2005
MODEL 1000
SETTING SOLENOID PURGE FLOWS
E-2
Have on hand the
sample system drawing
that was included with the GC system as part of the
shipping package. This drawing identifies, in schematic form, the location of solenoid purge
valves that you will need to operate during this procedure.
To set up solenoid purge flow, follow these steps
:
(1)
At the GC Controller, halt continuous analysis.
(2)
At the Analyzer, ensure that all streams to be analyzed are plumbed into the system and
stream pressures are set to 15-20 psig.
(3)
At the Analyzer, locate and
close
the solenoid purge valve for each solenoid. This is the
valve plumbed to the normally open (NO) port of each stream solenoid.
(a)
Refer to the sample system drawing for the location of these valves.
(4)
Access the Analyzer Valve Driver Board(s) with stream switches. Manually turn the switch
for stream 1 ON, and turn all other streams OFF.
(5)
Open the purge valve associated with stream 2, and regulate the flow through the solenoid
purge flow indicator to approximately 200 cubic centimeters per minute (ccm).
(a)
Refer to the sample system drawing for the location of purge valves.
(6)
Open the purge valve associated with stream 3, and regulate the flow through the solenoid
purge flow indicator to
add
approximately 200 ccm to current flow.
(a)
Flow indicator should now read 400 ccm.
(7)
Continue to open solenoid purge valves, adding 200 ccm flow for each additional stream (as
measured at the solenoid purge flow indicator).
Summary of Contents for Daniel Danalyzer 1000
Page 2: ......
Page 18: ...SEP 2005 MODEL 1000 DESCRIPTION 1 6 Figure 1 1 GC System Functional Block Diagram ...
Page 38: ...SEP 2005 MODEL 1000 EQUIPMENT DESCRIPTION 2 6 Figure 2 1 Analyzer Components ...
Page 146: ...SEP 2005 MODEL 1000 INSTALLATION AND SETUP 3 92 Figure 3 37 Analyzer Components ...
Page 188: ...SEP 2005 MODEL 1000 MAINTENANCE 5 20 Figure 5 2 Analyzer Components ...
Page 222: ...SEP 2005 MODEL 1000 MAINTENANCE 5 54 This page intentionally left blank ...
Page 228: ...SEP 2005 MODEL 1000 LIST OF APPENDICES A ii This page intentionally left blank ...
Page 260: ...SEP 2005 MODEL 1000 SETTING SOLENOID PURGE FLOWS E 4 This page intentionally left blank ...
Page 282: ...SEP 2005 MODEL 1000 ANALYZER DRAWINGS Addendum 1 2 This page intentionally left blank ...
Page 283: ......
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......
Page 311: ......
Page 312: ......
Page 313: ......
Page 314: ......
Page 316: ......
Page 318: ......
Page 319: ......