MODEL 1000
SEP 2005
MAINTENANCE
5-23
(6)
Plug the Sample Vent (labeled "SV") vent line.
(7)
Pressurize the calibration gas line to 50 psig.
Calibration gas line pressure of 50 psig is for leak check and test purposes only. For normal
operation, the calibration gas line pressure is maintained at 20-30 psig.
(a)
Shut the calibration gas bottle valve.
(b)
Observe the pressure on the high-side regulator gauge of the calibration gas bottle.
Because the "SV" vent line is plugged, the pressure should not decrease during a
period of 2-3 minutes.
(8)
This completes the calibration gas line leak check.
Next, leak check the sample gas lines,
according to the steps that follow
.
(9)
Plug the "SV" vent line (it may already be plugged if you performed steps (5) through (7),
above, to leak-check the calibration gas line).
(10)
Pressurize the sample gas line to 50 psig or a known pressure.
Sample gas line pressure of 50 psig is for leak check and test purposes only. For normal
operation, sample gas line pressure is maintained at 20-30 psig.
(a)
Shut off the sample gas.
(b)
Observe the pressure on any gauge that indicates pressure between the closed sample
gas block valve and the plugged "SV" vent line. Because the line is plugged, the
pressure should not decrease during a period of 2-3 minutes.
Summary of Contents for Daniel Danalyzer 1000
Page 2: ......
Page 18: ...SEP 2005 MODEL 1000 DESCRIPTION 1 6 Figure 1 1 GC System Functional Block Diagram ...
Page 38: ...SEP 2005 MODEL 1000 EQUIPMENT DESCRIPTION 2 6 Figure 2 1 Analyzer Components ...
Page 146: ...SEP 2005 MODEL 1000 INSTALLATION AND SETUP 3 92 Figure 3 37 Analyzer Components ...
Page 188: ...SEP 2005 MODEL 1000 MAINTENANCE 5 20 Figure 5 2 Analyzer Components ...
Page 222: ...SEP 2005 MODEL 1000 MAINTENANCE 5 54 This page intentionally left blank ...
Page 228: ...SEP 2005 MODEL 1000 LIST OF APPENDICES A ii This page intentionally left blank ...
Page 260: ...SEP 2005 MODEL 1000 SETTING SOLENOID PURGE FLOWS E 4 This page intentionally left blank ...
Page 282: ...SEP 2005 MODEL 1000 ANALYZER DRAWINGS Addendum 1 2 This page intentionally left blank ...
Page 283: ......
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......
Page 311: ......
Page 312: ......
Page 313: ......
Page 314: ......
Page 316: ......
Page 318: ......
Page 319: ......