SEP 2005
MODEL 1000
INSTALLATION AND SETUP
3-82
(a)
If necessary, see instructions in Section 3.4.1, step (3).
(2)
Locate the DB-25 female parallel printer port on the TB. It is labeled "P1".
(3)
Use a standard parallel printer cable (customer supplied) for making the connection between
the GC Controller and the printer's parallel port.
To connect a printer to one of the Controller's serial (COM) ports, follow these steps
:
(1)
Access the GC Controller's Terminal Board for Field Wiring (TB) ( See drawing DE-20782).
(a)
If necessary, see instructions provided in Section 3.4.1, step (1).
(2)
Choose an available serial port on the TB that is configured for RS-232 serial protocol.
(a)
Leave at least one serial port available for connecting the GC Controller to a PC
(usually COM1).
(b)
Also leave available the other serial ports planned for use by RS-232 serial devices.
(c)
See Section 3.4.4 for a complete listing of the TB's serial ports and their pinouts in
order to prepare a serial printer cable.
(3)
After the wiring connections have been completed, use the GC software MON2000 to
configure the GC Controller's serial port for use by a printer.
(a)
Select the "Application > Serial Ports menu path”.
(b)
Choose the serial port that correlates to the COM port chosen in step (2), above.
(c)
Set "Usage" to
Report
, "Prtcl" to
ASCII
and "RW" to
W
.
(d)
Set all other settings to default values (i.e., "Baud Rate"
9600
, "Data Bits"
7
, "Stop
Bits"
1
, "Parity"
None
, "RTS On"
0
, "RTS Off"
0
).
Summary of Contents for Daniel Danalyzer 1000
Page 2: ......
Page 18: ...SEP 2005 MODEL 1000 DESCRIPTION 1 6 Figure 1 1 GC System Functional Block Diagram ...
Page 38: ...SEP 2005 MODEL 1000 EQUIPMENT DESCRIPTION 2 6 Figure 2 1 Analyzer Components ...
Page 146: ...SEP 2005 MODEL 1000 INSTALLATION AND SETUP 3 92 Figure 3 37 Analyzer Components ...
Page 188: ...SEP 2005 MODEL 1000 MAINTENANCE 5 20 Figure 5 2 Analyzer Components ...
Page 222: ...SEP 2005 MODEL 1000 MAINTENANCE 5 54 This page intentionally left blank ...
Page 228: ...SEP 2005 MODEL 1000 LIST OF APPENDICES A ii This page intentionally left blank ...
Page 260: ...SEP 2005 MODEL 1000 SETTING SOLENOID PURGE FLOWS E 4 This page intentionally left blank ...
Page 282: ...SEP 2005 MODEL 1000 ANALYZER DRAWINGS Addendum 1 2 This page intentionally left blank ...
Page 283: ......
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ......
Page 288: ......
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......
Page 293: ......
Page 294: ......
Page 295: ......
Page 296: ......
Page 299: ......
Page 300: ......
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 309: ......
Page 310: ......
Page 311: ......
Page 312: ......
Page 313: ......
Page 314: ......
Page 316: ......
Page 318: ......
Page 319: ......