Doc# E151701 4-2
4: Performing Meter Testing
Electro
Industries/GaugeTech
™
Powered by Innovation™
4.1: Testing Tools
The Nexus® 1262 and 1272 meters are equipped with a light-emitting diode (LED) for
calibration and a liquid crystal display with test screens. These two standard features
provide user interface and equipment interface points for meter testing.
Figure 4.1: Meter Test Pulse and Test Button Location
4.1.1: Test Pulse LED
The Test Pulse LED is located in the upper left-hand portion of the meter face as
shown in Figure 4.1. The Test Pulse LED continuously emits calibration pulses (infra-
red light). This LED can be used for a variety of pulses including individual quadrants
for Watt-hour, VA-hour, VAR-hour and Watt Demand pulses. Each calibration pulse is
equal to the value assigned to Kt (Watt-hours per pulse) in the meter’s Device Profile.
The millisecond duration of each output pulse is also programmable in the meter’s
Device Profile.
To program the Test Pulse LED:
1. From the Device Profile menu in the CommunicatorPQA® application, click
Reve-
nue and Energy Settings>Internal KYZ
to configure the following settings:
Under Cover
Test Button
Infrared
Test Pulse