GBK60027 DX-SLM Installation Manual Issue 5
7.4
Calibrating the SLM using the HHP
Twelve SLM parameters must be calibrated to the individual wheelchair system using
the HHP. These parameters are all described in section 7.3.
The first four parameters calibrated by the HHP are approached sequentially. The
eight Restrictor Plate parameters are grouped and calibrated in one operation of the
HHP.
7.4.1
Initial Operation
1. Turn the DX System on and plug the HHP into the Programmer Socket on the DX
Remote. The initial screen appears for two seconds.
Dynamic DX
Programmer
VERSION 1.10
2. Then the main menu screen reads:
* * MAIN MENU * *
View or edit
System ?
YES
? DIAG
TECH
3. Press 'TECH' and a screen will appear to enter the three-digit password.
Technician Mode
Enter Password
0 0 0
EXIT
D1
D2
D3
Press the D1, D2 and D3 buttons to cycle each digit through to the correct
password. When the password reads correctly, press the 'EXIT' button.
4. The screen now reads:
Technical Mode
Master JS Module
JOYSTICK CALIBRATION
EXIT
YES
NEXT
5. Press 'NEXT" until the screen reads:
* * MAIN MENU * *
View or Edit Servo
Steering Module ?
NEXT
YES
Page 40
M
T