from the probe tip throughout the three calibration tests.
NOTE:
The front panel CALIBRATION INDICATOR blinks when the SCAN/Calibrate Control
is in either the BELOW ( - ), CENTERED (>0<), or ABOVE ( + ) test position. The
CALIBRATION INDICATOR is OFF when the CALIBRATION control is in the SCAN position.
NOTE:
All Dilon Technologies probes can be used with any Navigator 2.0 control unit.
The 12mm Lymphatic Mapping Probe, used with a Gain Module, requires a different
method of peak calibration assurance, entitled “Running a Peak Procedure".”
11C. Fuse Replacement Procedure
The fuse is to be replaced when necessary by the user (when the fuse is "blown"). The Navigator 2.0
fuse is to be replaced as follows:
1. Push-in and twist the fuse holder cap counter-clockwise.
2. Remove the fuse from the fuse holder.
3. Insert a new fuse into the holder.