Table 6C-1. Navigator 12 mm Probe - Settings and Indicators for use with Tc99 (just prior to
surgery)
Control/Indicator
Setting
Controls (back of Control Unit)
SCAN/Calibrate:
SCAN
Isotope:
Tc99
Indicators (front of the Control Unit)
Range:
1x
Threshold:
Illuminated
Display:
0
Isotope:
Tc99
Gain Module
Dial at Peak Setting:
Perform peak procedure
NOTE:
Keep control unit power off, until all components are connected.
NOTE:
A ‘peak procedure’ must be performed before using the probe in the first surgical procedure
of the day
.
NOTE:
Although the ‘peak procedure’ is typically performed with no sterile drape around the probe
and cable, it may also be performed with the probe and cable inside a sterile drape.
NOTE:
After a ‘peak procedure’ has been performed, the control unit and gain module settings are
given in the table above.
NOTE:
Follow the instructions on ‘Cleaning, Disinfection, and Sterile Use of Probe and Cable’ (Dilon
doc. # 141-00005).
CAUTION:
For intraoperative use, insert probe and cable in a sterile drape.
6C-2. During Surgery
·
See
‘
Useful Adjustments that can be made During Procedures’
.
6C-3. After Surgery
·
See:
‘
Cleaning, Disinfection, and Sterile Use of Probes and Probe Cables’ (Dilon doc. # 141-
00005).