Table 6A-1. Pilot Probe LED Indicator
Indication
Status
On/Flashing
Probe is linked and ready for use.
Off
Probe is in a resting position to conserve power when the Control Unit is off; to
reactivate LED indicator, turn on the Control Unit and simply pick up probe, or if
needed, lightly shake probe
If no power upon ready to use, the battery needs
to be installed or replaced.
If battery has been replaced and LED light is still off, contact your
distributor or Dilon Technologies directly.
6A-2. During Surgery
·
See
‘
Useful Adjustments that can be made During Procedures’
.
NOTE:
For Technetium-99m (Tc99), the control unit settings are given in the following table.
NOTE:
Follow the instructions on ‘Cleaning, Disinfection, and Sterile Use of Probe’ (Dilon doc. # 141-
00005).
Table 6A-2. Navigator 2.0 with Wireless Pilot Probe – Settings & Indicators (just prior to
surgery)
Control / Indicator
Setting
Controls (back of Control Unit)
SCAN/Calibrate:
SCAN
Isotope:
Tc99
Indicators (front of the Control Unit)
Range:
1x
Threshold:
Illuminated
Display:
0
Isotope:
Tc99
Indicator (top of the Pilot Probe handle)
Probe LED:
Illuminated; flashing.
6A-3. After Surgery
·
See:
‘
Cleaning, Disinfection, and Sterile Use of Navigator Probes & Cables’ (Dilon doc. #
141-00005).
6B. Navigator 2.0 with Standard Lymphatic Mapping Probes, Superficial Head & Neck