5. Probe Connectivity and Use
6A. Navigator 2.0 with Wireless Pilot Probe
The Wireless Pilot Probe is used in various procedures. A typical sequence of setting up the Wireless
Pilot Probe for a procedure with a Technetium-99m isotope (such that may be used in a lymphatic
mapping procedure for a sentinel node biopsy) is as follows:
6A-1. Before Surgery
·
Insert a charged battery into control unit.
·
Upon initial insertion of new Wireless Pilot Probe battery, probe may need to be lightly shaken
to activate LED in probe base.
·
LED on the Pilot Probe indicates that it is linked with control unit and ready for use. When the
Control Unit is turned off, the Probe
’s LED will turn off within seconds of being placed in a
resting position. When the Control Unit is turned on, the Pilot Probe will instantly power up
when moved. It is now ready for use.
·
For intraoperative use, insert the Wireless Pilot Probe into a sterile drape.
·
Insert a probe battery into the Pilot Probe as follows:
1. Hold probe firm; turn battery cap counterclockwise and remove from probe. Inspect O-
ring integrity. If O-ring is missing or damaged, use new battery cap. Contact Dilon
Technologies or your distributor for battery cap reorder information.