Navigator 2.0 User Manual & Service Guide
N2-5000-07-001 Rev 7
-
21-
Revised 10/16/2018
www.Dilon.com
4D. Optional Co-Pilot™ Device (Part # GP-6801-00)
The optional Co-Pilot is a single-use device used for initiating counting periods and adjusting the
audible range from the probe, inside the sterile field. It includes two small buttons, and a long, small-
diameter cable.
To plug in the Co-Pilot, match the spacing of the prongs with the spacing of the receptacles on the
control unit. It should then be clipped onto the base of the probe.
The Co-Pilot is supplied sterile and may be used inside or outside of the sterile drape.
The
“C” button is the COUNT Button. Obtain a one-second count by pushing and releasing this button
once. Obtain a ten-second count by pushing this button twice, in quick succession. Each time, total
counts are shown in the display screen on the control unit.
The
“R” button is the RANGE Button. This button operates the Range control mentioned above on
the control unit. Push and release the RANGE Button to select an audible range, appropriate to the
signal detected by the system.
CAUTION:
The Co-Pilot can only be attached to the control unit one way – the prongs are not
evenly spaced (see image below).
C Button (Count)
R Button (Range)