Navigator 2.0 User Manual & Service Guide
N2-5000-07-001 Rev 7
-
16-
Revised 10/16/2018
www.Dilon.com
Table 4A-2. Controls and Displays on the Back of the Control Unit
Control Display
Description
For CABLED PROBES only
. The SCAN position is the only
correct position when a probe is being used during a procedure.
When set to SCAN, the CAL indicator on the front of the control
unit will not illuminate.
For instructions on use of the
‘+’, ‘0’, and ‘-‘positions, see
“Verification of Standard Gain (Calibration Quick Test)”.
Selects the isotope to be detected by the control unit.
SCAN/Calibrate Control
For CABLED PROBES only. The SCAN/Calibrate Control has FOUR POSITIONS. During all
surgical procedures, this control should be in the SCAN Position. During the calibration
verification procedure, this control uses the remaining three settings, “+”, “0”, and “-”.
NOTE
:
See “Verification of Standard Gain (Calibration Quick Test)” for information on calibration
verification.
NOTE:
If the front panel CAL indicator is flashing before a procedure, move the control to the SCAN
position.
Isotope Control
The Isotope Controls allow the user to designate the specific isotope in use.
Switch set on: I125
Switch set on:
511keV
Switch set on: In111
Switch set on: Tc99
Iodine-125
18
F-FDG
(and I131)
Indium111
Technetium-99m
The Isotope Control setting on the back of the control unit illuminates the corresponding light on the
Isotope Indicator on the front of the control unit.
CAUTION:
It is important that the isotope control is set to the isotope that is going to be used in
the procedure. Setting the isotope control incorrectly will result in incorrect detection.