Navigator 2.0 User Manual & Service Guide
N2-5000-07-001 Rev 7
-
22-
Revised 10/16/2018
www.Dilon.com
4E. Useful Adjustments That Can Be Made During Procedures
Table 4D-1. Useful Adjustments
Adjustment
Benefit
Threshold
For Cabled Probes only
, this feature increases specificity when only a low
number of events are observed. Threshold control defaults to ON. When ON, the
system counts only the events in a narrow energy range around the signal.
Change Threshold to OFF to allow the system to count all signals it detects,
opening the value range to scatter. (NOTE: The Wireless Pilot Probe features
integrated threshold.)
Range
The Range function defaults to 1X, meaning that the audible signal fluctuates
according to every single count it detects. In the 10X position, every 10
th
event
produces an audible output. In 100X, only every 100
th
event produces an audible
output.
The Range control only affects the sound. The count shown in the display is
independent of the range setting.
10-Second
Count
Press to obtain a 10-second count, keeping probe in fixed position each time.
The total is displayed for at least four seconds, allowing time to record the total.
Volume
Adjust to desired volume.
Power
Press to turn on the control unit or to safely turn off the device.
10-second Count Button and
Indicator
Range Button and
Indicators
Volume Knob
Threshold Control
Power Button