Navigator 2.0 User Manual & Service Guide
N2-5000-07-001 Rev 7
-
33-
Revised 10/16/2018
www.Dilon.com
Problem
Possible Causes
Remedies
2. LED on Pilot
Probe does not
illuminate.
Probe battery is dead or
installed incorrectly.
Replace with new battery.
Verify that battery was inserted
correctly into probe (
‘+’ should face
toward base of probe).
Battery was not installed.
Install new battery.
LED on probe is damaged.
Contact Dilon Technologies for
assistance.
Table 8A-3. Control Unit with CABLED PROBE- Settings and Indicators
Problem
Possible Causes
Remedies
1. Zero in display.
No signal under
presence of a
radioactive source.
No connection between
probe, cable, gain module (if
present), and control unit.
Check that all connections are secure.
Isotope control is set to
incorrect isotope.
Change Isotope Control (on back of
control unit) to Tc-99.
Gain module, if used, is set
to zero.
Run Peak Procedure (Sec. 7A).
There is an open circuit in
the probe cable.
Replace cable.
Circuit inside the control unit
has been damaged.
Try a different control unit. Contact
Dilon Technologies for assistance.
Probe is damaged / probe
cap not secure / detector
damaged.
Try a different probe or contact Dilon
Technologies for assistance.
2. Spurious high
counts, such as
80,000 counts a
second (when probe
is held in air, for
example).
Intermittent short in the
cable.
Replace cable.