Navigator 2.0 User Manual & Service Guide
N2-5000-07-001 Rev 7
-
31-
Revised 10/16/2018
www.Dilon.com
during one surgical day with the same control unit
1. Probe A
is going to be used on Monday for all surgical procedures scheduled on
that day.
2.
Peak Procedure is run on
Probe A
before the first surgical procedure of that day.
3. Probe A
counts are highest when the Gain Module dial is set to about 4.5. The dial
is left at that location.
4. Probe A
is ready for all surgical cases for the day.
5. Probe A
is dropped and damaged. It cannot be used again until tested and/or
repaired.
6.
The Surgical Team chooses to use
Probe B
for the rest of the case and for all
others scheduled on that day.
7.
A Peak Procedure is run on
Probe B
before the next case or before continuing the
current surgical procedure to
“pair” it with the control unit.
8. Probe B
counts are highest when the Gain Module dial is set just above 6. The dial
is left at that location.
9. Probe B
is ready for all surgical cases for the day.