Navigator 2.0 User Manual & Service Guide
N2-5000-07-001 Rev 7
-
14-
Revised 10/16/2018
www.Dilon.com
Control Display
Description
Count
: Initiates a 10-second photon count.
When Count has been pressed, the count indicator on
the display screen is illuminated and the display screen
will show increasing counts. Probe must be held in a
fixed position for entire duration of 10-second count.
When the 10 seconds are complete, the control unit
beeps, and the total count is shown in the display.
After displaying the total count for four seconds, the
display goes back to showing counts per second.
Cal-check:
This light indicates when the system is in
‘Calibration-Check’ mode on the back of the unit. The light will
be illuminated when in any of the 3 calibration-check settings,
and will not be illuminated when the system is set to the
‘Scan’ mode.
The
SCAN / Calibrate Control must be set to the SCAN
position only, for all probes for all procedures. In this
mode, the ‘Cal’ light will be turned off.
See
“Verification of Standard Gain (Calibration Quick Test)”
for more information on Calibration.
The
Battery indicator
shows the charge status of the battery
in use.
When the indicator level on the control unit is at 25%, the
battery should be replaced immediately with a fully-charged
battery.
Please note that the charge status on the control unit may
differ from charge status reflected on the battery, due to a
higher power requirement on the control unit. Refer to charge
status on control unit rather than the battery itself.
Signal input port
, for cable connection. The signal input port
is
not
applicable when using the Wireless Pilot Probe with the
Navigator 2.0.
For the 12mm Lymphatic Mapping Probe, connect the cable
attached to the Gain Module here, matching the arrows on
the cable connector to the arrow above the signal input port.
See
“3mm Diameter Cable” for more information.
For all other cabled probes, connect the probe cable here,
matching the arrow on the cable connector to the arrow
above the signal input port. See
“6mm Diameter Cable” for
more information.
Connection port for the optional
Co-Pilot
accessory.
See
“Optional Co-Pilot Device” for more information.