Navigator 2.0 User Manual & Service Guide
N2-5000-07-001 Rev 7
-
18-
Revised 10/16/2018
www.Dilon.com
4B-3. Charging the Battery (Battery Charger Part # N2-8000-02)
Place the charger on a flat, level surface, away from sources of heat and moisture. Plug the DC
connector from the power supply into the back of the charger, and connect the power supply to the
mains AC, using the cable supplied. All of the LEDs will flash momentarily to indicate that power is
present.
Recharge Time: Approximately 3.5 hours
Place the battery into either battery bay, ensuring that the 5-way connector is fully seated. The battery
should feel secure once inserted correctly. The LEDs in the battery status window will provide charge
status, and the charger will automatically begin charging. Each charge bay operates independently,
providing simultaneous charge of each battery inserted.
While there is a calibration button for each bay,
calibration is not necessary for use with the
Navigator 2.0.
If calibration button is inadvertently pressed, either a flashing blue or solid blue light
will illuminate. Simply remove battery and reinsert in order to resume charge. A green light will indicate
that it is in charge mode.
If charger has flashing red light, the battery fuel gauge requires calibration. Only in this case, should
the calibration button be pressed. Recalibration can take 10-13 hours.
If charger lights solid red, please refer to troubleshooting guide.
Table 4B-4. Charge Bay LED Indications
Indication
Battery Charge Status
Green Flashing
Battery Charging
Green Solid
Battery Fully Charged
Blue Flashing or Solid
Calibration mode: UNNECESSARY. Do not use.
Red Flashing
Battery fuel gauge in need of calibration
Red Solid
Error
NOTE:
Use only batteries supplied by Dilon Technologies. The Dilon Technologies control unit
battery has the proper dimensions and a key feature that holds it securely in the Navigator
2.0 control unit.