7.3
‘Check Probe’ Error Message
This message typically indicates an error generated by the
probe tip being wet. Dry the tip and cycle the device power
off then back on. If drying the tip of the probe does not
resolve the error, then the probe may have degraded to the
point that it will require replacement.
7.4
‘Plug In Probe’ Error Message
This message occurs when: (1) the detachable probe is not
mated or is improperly mated to the unit, or (2) the probe is
defective. If the probe is found to be defective, remove
defective probe by holding the probe connector and gently
pulling straight out of the unit.
WARNING:
Do not twist probe as this could damage connectors. Properly align the probe
connector and gently push in until properly seated.
7.5
‘PQF Failed’ Error Message
This message usually indicates a hardware failure occurred
within the unit and the unit must be returned for repair.
See section 19.1 for service information.
CHECK
PROBE
PLUG IN
PROBE
PQF
FAILED
!