Page 39 of 87
DGH 55B-INS-OMENG Rev: 3
12.2.14
If a questionable measurement is found during review, the operator can delete it. To do
this, the operator presses the DEL key while viewing the measurement in question. The
Measurement Average and Standard Deviation will be automatically updated.
•
The operator can take new measurements to replace those that were deleted or
choose to accept the remaining measurements.
12.2.15
Once both the OD and OS measurement groups are complete, no more measurements
can be taken for that eye unless the measurements of that group are cleared. (Or
individual measurements are deleted, as described in 12.2.14)
12.2.16
If the device is operating in Multi-Patient Mode the measurements will be automatically
saved to memory.
12.2.17
To clear all measurements for one or both eyes,
press the CLR key. The user will be prompted with
the ‘What To Clear?’ menu:
•
The following selections affect the patient currently being measured.
o
Selecting ‘All Current Meas’ clears all measurements of both the left and
right eye for the currently loaded patient.
o
Selecting ‘OD Current Meas’ clears all measurements of the right eye for
the currently loaded patient.
o
Selecting ‘OS Current Meas’ clears all measurements of the left eye for
the currently loaded patient.
o
Selecting ‘Nothing (Exit)’ exits the menu without deleting anything.
•
The other clearing options affect stored patient information or system configuration.
o
Selecting ‘All Patients’ clears all identifying information and all
measurements for all patients. (Bluetooth
®
option only)
o
Selecting ‘Paired Devices’ clears all paired devices from the device
memory. (Bluetooth
®
option only)
WHAT TO CLEAR?
All Current Meas