6 Overview of modules and boards
Service Manual Sinius / Sinius CS / Sinius TS
6.6 LEDview
båÖäáëÜ
6.6
LEDview
6.6.1
NLV board - LEDview control
Components on the board
LEDs
V201
24-32V DC
(green)
32V input voltage
V209
3.3 V DC
(green)
3.3V operating voltage
V215
5V DC
(green)
5V operating voltage
V301
RUN
(green)
Flashes during standard operation
V303
RESET
(red)
Lights up briefly after being switched on
V305
BOOT
(yellow)
Lights up after switching on and remains lit until completion
of the boot process (approximately 60 seconds)
If the board remains in the bootloader after this:
● No launch command has been received from the
download manager, as the software is not consistent
● There is no connection to the HSA.
V307
ERROR
(red)
Flash code 2x: There is no connection to the HSA.
Connector
X1
> LEDview cable > X1 > L225 > X28 NWEm