10 Biofilm removal
Dentsply Sirona
10.3 Execution
Service Manual Sinius / Sinius CS / Sinius TS
10.3
Execution
Preparation / cleaning solution 1
1. Mix cleaning solution 1 in accordance with the manufacturer’s
instructions.
2. Pour all the solution into the biofilm removing system tank.
3. Switch the treatment center and the biofilm removal system pump on.
Rinsing / cleaning solution 1
1. Rinse all instrument hoses carrying water for approx. 10 seconds by
activating the relevant instrument multiple times. Switching the water
flow on and off supports the process for removing deposits.
2. Activate the tumbler filler 1x.
3. Activate the flushing 1x.
4. Activate the suction hose cleaning 1x (if present).
5. Rinse the external device connection (if present) by connecting the
tool described above and opening and closing its valve multiple times
until cleaning solution flows out of the hose.
Switching the water flow on and off supports the process for removing
deposits.
Close the valve again.
6. Let cleaning solution 1 take effect in accordance with the instructions
from the manufacturer of the cleaning agent.
7. Complete steps 1-6 until the tanks are empty of all solution. Ensure
that the mixing tank is also empty.
8. Switch off the pump on the biofilm removing system.
Preparation / cleaning solution 2
1. Mix cleaning solution 2 in accordance with the manufacturer’s
instructions.
2. Pour all the solution into the biofilm removing system tank.
3. Switch on the pump on the biofilm removing system.
WARNING
Wear safety gloves, safety goggles and mouth and nose protection
when performing the following work.
IMPORTANT
Collect the colored cleaning liquid as applicable in a suitable container
for subsequent inspection.
IMPORTANT
The solution in the collection container must be clear following the
rinsing process. If the solution is still cloudy, fill the tanks with a new
cleaning solution 1 again and complete cleaning steps once again. The
cleaning liquid can then be removed via the cuspidor.