6 Overview of modules and boards
Dentsply Sirona
6.3 Dentist element
Service Manual Sinius / Sinius CS / Sinius TS
V304
Reset
(red)
Lights up shortly after switching on.
V102
Chair stop
(yellow)
V302
BL
(yellow)
Lights up after switching on and remains lit until completion
of the boot process (approximately 60 seconds)
If the board remains in the bootloader after this:
● No launch command has been received from the
download manager, as the software is not consistent
● There is no connection to the HSA.
V305
User
(yellow)
Lights up when at least one instrument has been removed.
V405
NaCl
(yellow)
ON: NaCl pump switched on
V600
FPGA
(yellow)
Flashing: OK (configuration is loaded)
Lit up: US scaler is active
V719
P5V
(green)
SPRAYVIT supply
Only when SPRAYVIT has been removed
Connector
X1
> L432 > X1 HAP2
X2
> Sirosonic
X3
> L451 > scaler
X5
> Instrument holder 1 Sprayvit with light, as well as air and water heater
X6
> L256 > Instrument holder 1 Sprayvit with light, keys, and hose coding
X7
> Spray water heater
X8
> L244 > X8 NAC/HAC
X12
> L260 > X2 CC
X19
> MV_21_1 - MV_21_4, MV_22_ZEG_water, MV_24_Sprayvit_water
X21
> L423 > SDI X1.12, X1.14 > L401b > X30 HSA
X22
> L423 > SDI X1.1, X1.3-X1.7, X1.9-X1.10 > L401a > X1 HSA
X24
> NaCl
X31
> L257 > X31 NAA
Components on the board