![Dentsply Sirona SINIUS Service Manual Download Page 68](http://html1.mh-extra.com/html/dentsply-sirona/sinius/sinius_service-manual_2483702068.webp)
63 69 149 D3561
65
D3561
.
076.01.07
.
02 02.2017
6 Overview of modules and boards
Dentsply Sirona
6.3 Dentist element
Service Manual Sinius / Sinius CS / Sinius TS
V311
Error
(red)
● 2 x No connection to HSA
● 3 x Self-test failed.
● 4 x Defective FRAM data.
● 5 x Critical error - reboot required.
● 6 x Error in motor behavior, motor does not start.
● 7 x Error in hardware foot control signal.
● 8 x Error in instrument detection.
● 9 x Inconsistent parameters received.
● 11 x Internal communication problem.
● 12 x Programming errors.
V301
Run
(yellow)
Flashes at 1Hz in standby
Flashes at 10Hz when an instrument is active
V302
Test 1
(yellow)
Lights up when the foot control is pressed.
V303
Test 2
(yellow)
Messages from the J416 motor controller
Flashes 5x: Programming
Flashes 6x: Error
Flashes 7x: Warning
V304
Boot
(yellow)
Lights up after switching on and remains lit until completion
of the boot process (approximately 60 seconds)
If the board remains in the bootloader after this:
● No launch command has been received from the
download manager, as the software is not consistent
● There is no connection to the HSA.
V312
Claw active
(yellow)
Displays the active instrument holder via flash code.
V901
5V DC
(yellow)
Operating voltage
Connector
X1
> L248 > X1.16 / X1.17 AK
X3
> L251 > Flange_scaler
X5
> Option_PK_supply
X6
> Option_PK_communication
X8
> > L244 > X8 NAJm
X12
> L258 > Instrument holder 2
X13
> L258 > Instrument holder 3
X14
> L258 > Instrument holder 4
X15
> L258 > Instrument holder 5
Components on the board