7 Replacing components
Service Manual Sinius / Sinius CS / Sinius TS
7.4 Hydraulic/pneumatic modules
båÖäáëÜ
7.4.3
Mixing tank
The mixing tank needs to be fully emptied before being replaced.
✔
The treatment center is switched on.
1. Hold down the "Setup" fixed key on the user interface of the dentist
element for at least 2 seconds.
Six setup dialogs are offered for selection.
2. Touch the "Service area" key for at least 2 seconds.
The service area will appear.
3. Use the arrow key to scroll to the next screen.
4. Touch the "Empty the mixing tank" key.
The mixing tank is emptied through the tumbler outlet.
5. Wait until the mixing tank is empty.
6. Turn the treatment center OFF at the power switch.