5 Important information
Dentsply Sirona
5.2 Switching the treatment center on/off
Service Manual Sinius / Sinius CS / Sinius TS
Third phase (approximately 40 seconds)
● Once the start release has been issued by the HSA, the software
boots in the corresponding nodes.
● All BOOT LEDs = OFF (yellow)
● All RUN LEDs flash (green)
● The tooth is displayed in the user interface.
● The BOOT LED remains ON for any nodes not released.
● HSA saves all available nodes. Should a node be missing since the
last switch-on, a service code is issued.
HSA checks the validity of the referencing data in the nodes HSA and
NSK, and issues corresponding movement releases (see
Referencing the drives [ → 35]).
● All nodes save the options detected in the configuration log. If an
option is no longer detected following a restart, a service code is
issued.
HSA saves all service codes reported in the system.