![Dentsply Sirona SINIUS Service Manual Download Page 25](http://html1.mh-extra.com/html/dentsply-sirona/sinius/sinius_service-manual_2483702025.webp)
63 69 149 D3561
D3561
.
076.01.07
.
02 02.2017
22
Dentsply Sirona
5 Important information
Service Manual Sinius / Sinius CS / Sinius TS
5.2 Switching the treatment center on/off
båÖäáëÜ
5.2.3
Signal path of the ON/OFF signal
NAJm
SDI
HSA
Power
supply
NWEm
Unit
ON/OFF
Unit ON/OFF
Powersupply
ON/OFF
Unit_ON
Yellow
X1
VT
OG
OG
BN
OG
L432
L423
L401a
L207
L400a
X22
X2
X1
X1
X1 X1
X1
X1
X5
B5
B5
10
10 10
B1
B1
10
2
B1
Logik
HAP2
SDI
G
O
23
L40
X1 X1
10 10
inexistent for
SINIUS CS
Unit_ON/OFF:
(keyboard up to NWEm)
Power supply unit_ON/OFF:
(NWEm to the power supply)
Switching on
● Approximately 8-10 V when the ON/OFF
key is not pressed
● 0 V when the ON/OFF key is pressed
● Approximately 8-10 V when the ON/OFF
key is not pressed
● Approximately 0.5 V when the ON/OFF key
is pressed
Switching off
● Approximately 28 V when the ON/OFF key
is not pressed
● Approximately 0.1 V when the ON/OFF key
is pressed
● Approximately 8-10 V when the ON/OFF
key is not pressed
● Approximately 0.5 V when the ON/OFF key
is pressed