8.2. Treatment Modes
Set the desired exposure mode using the «Mode Select» bu on switch.
The
desired mode is selected by pressing the «Mode Select» bu on (Mode 1 – Mode
2 – Mode 3 – Mode 1) in sequence, and a sound signal is heard repeatedly where
the number of repe ons corresponds to the mode number. Mode indica on is
provided by corresponding LED indicator.
The desired treatment mode is set with «Mode Select» bu on as indicated in
Table 1.
Table 1
Operating
mode
Field
type
Inductor coil
excitation
frequency
Peak (amplitude) value of
the magnetic field density
on the inductor coil
working surface
Mode
description
mT
G
1
traveling
6.25
(1/8 of the mains
frequency 50 Hz)
20±6
200±60
Basic mode
of operation
2
traveling
6.25
(1/8 of the mains
frequency 50 Hz)
8±2
80±20
Pediatric mode
with reduced
magnetic field
density
3
stationary
100
(doubled mains
frequency 50 Hz)
6±2
60±20
Therapeutic mode
with pronounced
analgesic and
anti-inflammatory
effects