OPERATING MANUAL
ENG
19
A en on!
A en on!
A en on!
Before star ng the procedure, de
fi
ne the con
fi
gura ons of inductor coils in the
form of a «
fl
exible spline» or «matrix» (depending on the requirements of the
methodology for trea ng a speci
fi
c disease).
The traveling pulsed magne c
fi
eld is to be directed from the 1
st
inductor coil to
the 4
th
inductor coil when «spline» emi er con
fi
gura on is used, namely:
The coil directly connected to the control unit is referred to as the 1
st
coil. Where
the matrix emi er con
fi
gura on is used, the inductor coils queuing and the direc-
on of the traveling pulsed magne c
fi
eld are similar.
In case the exposure results in the blood pressure increase or de-
crease by 10 to 25 mm Hg during the
fi
rst 6 days of treatment
(Mode 3), reduce the treatment dura on by 1/3 during the next ses-
sion and apply Mode 2 during the next 3 days of treatment. From
that me onwards, resume Mode 1 treatment.
In case the exposure results in the blood pressure increase or de-
crease by more than 25 mm Hg consult with your a ending doctor
prior to the next session to adjust the treatment procedure. Keep a
log by entering the measurement data required to trace the me
course of blood pressure.
!
!
It is not recommended at the same me (on the same day) to treat
two di
ff
erent diseases.
!