8 Software Download and Diagnostics
1505-1018-000 08/07
8-5
( P a g e 2 o f 2 )
P o w e r C o n t r o l P o w e r D i a g n o s t i c s
Note: Fan 1 is located on the rear cage, Fan 2 is located on the PMB.
Turn the ComWheel to return to the first page.
Push the ComWheel to return focus to the Power Diagnostics selection menu.
Battery Status
Power Diagnostics
Power Control
Display Unit
->Main Menu
Label
Value Format
Units
Normal range
Board Temperature
<65C OK (green)
>65C <75C Warn (Yellow)
>75C Fail (red)
Deg C
Upper Limit 64
Fan Speed
Slow, Fast
Fan 1 Voltage
XX.XX
(high fan speed = 23.7V typical)
(low fan speed = 20.7V typical)
Vdc
17.5 to 25.0
Fan 1 Status
Fail, OK
Fan 2 Status
Fail, OK
Battery Status
Description
Fail
Battery is disconnected from Power Control Power
Diagnostics and is in Trickle mode for > 12 hours.
Bulk Chg
Initial phase of battery charging. Battery is at 600 mA
when system is running and 1.2 Amps when system is in
standby.
Over Chg
Topping off remaining 5-10% of battery capacity and
approaching full charge.
Float Chg
Battery is being maintained at full charge.
Trickle Chg
Independent of battery charger. Severely degraded
batteries are charged at 25 mA constant current when
either battery is 6-10 Vdc and batteries are not
connected to Power Control Power Diagnostics.
Discharge
System is running on battery power.
Summary of Contents for Engstrom Carestation
Page 1: ...Engström Carestation Technical Reference Manual ...
Page 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Page 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Page 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Page 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Page 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...