9 Repair Procedures
1505-1018-000 08/07
9-23
9.5.5 To replace
the LCD display
1. Disconnect the following cables:
• Right membrane switch flex-cable at ZIF (zero insertion force) connector (
A
)
• Encoder assembly cable (
B
)
• Lower membrane switch flex-cable at ZIF (zero insertion force) connector (
C
)
• Fan cable (
D
)
• LCD cable (
E
)
• Left membrane switch flex-cable at ZIF (zero insertion force) connector (
F
)
2. Remove the ten screws (
circled
) that hold the mounting plate to the front enclosure.
3. Remove the mounting plate assembly from the front enclosure.
4. Disconnect the backlight harnesses (
G
) from the inverter boards.
5. Slide the grommet (
H
) out of the mounting plate slot (transfer to new LCD).
A
B
D
E
C
F