Engström Carestation
5-18
08/07 1505-1018-000
5.2.9 Verify
operation of free-
breathing valve
1. Ensure both manual shut-off valves are closed.
2. Connect a negative pressure squeeze bulb to the
INSP
iratory outlet.
3. Fully depress the bulb a minimum of 10 times.
4. On the VCB screen:
• Verify that the measured
Inspiratory Press
ure is more positive
than - 3 cmH
2
O at all times.
• if not, troubleshoot the free-breathing check valve and the inspiratory effort valve.
5. Disconnect the squeeze bulb from the
INSP
iratory outlet.
AB.98.091