8 Software Download and Diagnostics
1505-1018-000 08/07
8-25
8.7.4 Baro P cal
8.7.5 Acutronic
8.7.6 EEProm
EEProm is used by manufacturing only.
Item
Description
Enter Pressure
Enter the local barometric pressure in mmHg. The Baro P
Calibration will use this value to calibrate the barometric pressure
transducer.
Baro P Cal
Starts the barometric pressure transducer calibration. The
calibration will Fail if nothing is entered in the Enter Pressure box.
Status
Displays the results of the barometric calibration (Passed or
Failed).
Item
Description
SW Ver
Software version of the acutronic Sensor software.
Exp Zero ADC
Displays the Analog to Digital Counter zero value fro the
expiratory flow sensor.
Exp Zero DAC
Displays the Digital to Analog Counter used to zero the
expiratory flow sensor.