4 Installation and Service Menus
1505-1018-000 08/07
4-7
4.2.3 Ventilator Settings
4.2.4 Parameter Settings
Menu Item
Message text
Values
Timing
Change timing setting: I:E, Tinsp, or
Tpause. Tpause is only available
when Flow is set to On.
I:E, Tinsp, Tpause
Flow
Select to use Flow settings or Insp
Pause. Tpause Timing is available
when Flow is set to On.
On, Off
Modes with
Backup
Select Modes that will allow backup
ventilation.
SIMV-VC (Yes, No)
SIMV-PC (Yes, No)
Bi-Level (Yes, No)
CPAP/PSV (Yes, No)
SIMV/PCVG (Yes, No)
Previous Menu
Menu Item
Message text
Values
TV Based on
Change value calculations: ATPD (no
humidifier) or BTPS (heated
humidifier).
ATPD, BTPS
CO2 Numbers
Change humidity compensation type
in CO2 partial pressure values.
Dry, Wet
Previous Menu
Return to previous menu.