3 Checkout Procedure
1505-1018-000 08/07
3-3
5. Select
Previous Menu
.
Checkout includes the following checks:
•
Paw Transducer Check
•
Barometric Pressure Check
•
Relief Valve Check
•
Exhalation Valve Check
•
Expiratory Flow Sensor Check
•
Air Flow Sensor Check
•
O2 Flow Sensor Check
•
O2 Concentration Sensor Check
•
Neonatal Flow Sensor Check (if applicable)
•
Circuit Leak, Compliance, and Resistance
3.3 Backlight test
1. Access the Calibration menu.
• In the standby mode, push the
System Setup
key.
• On the System Setup menu, select
Install/Service
(23-17-21).
• On the Install/Service menu, select
Calibration
.
2. On the Calibration menu, select
Backlight Test
.
3. Select
Start Test
.
4. The display will show the test running on light 1 and then on light 2. If the
display goes completely blank or flickers during the test, one of the lights
has failed.
3.4 Power failure test
1. Connect the power cord to a wall outlet. The mains indicator on the front
panel of the Display Unit comes on when AC Power is connected.
2. Set the system switch to On and Start a case.
3. Unplug the power cord with the system turned on.
4. Make sure the following message is displayed:
• ‘On battery’
5. Connect the power cable again.
6. Make sure the alarm cancels.
Summary of Contents for Engstrom Carestation
Page 1: ...Engström Carestation Technical Reference Manual ...
Page 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Page 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Page 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Page 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Page 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...