Engström Carestation
7-26
08/07 1505-1018-000
No battery
backup?
(Note "?" in
alarm message)
DU to PMB Com Error
-or-
PMB to DU Com Error
Medium
Communications from the DU to the
power controller cannot be established
for 10 seconds.
-or-
Communications from the power
controller to the DC cannot be
established for 10 seconds.
PMB assumes failed state and
allows system to shutdown
without confirmation from DU.
(This alarm allows vent to be
shut down if DU
communications is lost.)
Note question mark at end of
message and no battery
status indicator on display.
• Check error logs for related messages.
• Check communication (no software information indicates no communication).
• If multiple failures, replace DU CPU board.
• Check / replace display cable.
• Check (reseat) all connections on all PC boards.
• Check/replace DU interface board.
• Replace motherboard.
• Replace PMB.
No battery
backup?
(Note "?"in
alarm message)
PMB POST Failure
Medium
Power controller failed CPU self tests
Note question mark at end of
message and no battery
status indicator on display
• Cycle power.
• Replace PMB.
No battery
backup
Battery Charge Fail
-or-
Standby Current High
Medium
The system is powered on with a
battery current >1.3 amps.
-or-
System is in Standby and
the charge current is >1.7 Amps.
Refer to Section 8.1.3 for more
information.
• Cycle power.
• Check/replace harness.
• Replace Power Supply.
• Replace PMB.
No battery
backup
Battery Fail
Medium
While bulk, over, or float charging any
battery is <10.5VDC (short)
-or-
Battery has been charging for >24h
while powered on (sulfated)
-or-
Voltage >16.5V during bulk or over
charging and normal current >0.25
amps (sulfated).
Battery status indicator
appears under clock.
Refer to Section 8.1.3 for more
information.
• Cycle power.
• Check/replace harness.
• Replace batteries.
Alarm message
Alarm ID
Priority
Alarm Condition
Special Behavior/Comments
• Action/Troubleshooting
Summary of Contents for Engstrom Carestation
Page 1: ...Engström Carestation Technical Reference Manual ...
Page 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Page 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Page 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Page 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Page 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...