Engström Carestation
8-4
08/07 1505-1018-000
8.1.3 Power controller
power diagnostics
There are two pages of diagnostics for the Power Controller.
Selecting
Power Control
brings up the first page of the Power Controller
Diagnostics.
( P a g e 1 o f 2 )
P o w e r C o n t r o l P o w e r D i a g n o s t i c s
Turn the ComWheel to select the second page.
Push the ComWheel to return focus to the Power Diagnostics selection menu.
Power Diagnostics
Power Control
Display Unit
->Main Menu
Label
Value Format
Units
Normal range
AC Status
OK, Fail
12Vdc Supply
XX.XX
Vdc
11.70 to 12.30
3.3Vdc Supply
X.XXX
Vdc
3.201 to 3.399
1.5 Vdc Supply
X.XXX
Vdc
1.45 to 1.55
Battery Connected
Yes, No
Battery Status
Fail,
Bulk Chg,
Over Chg,
Float Chg,
Trickle Chg,
Discharge
Battery Current
X.XXX
A
Battery 1 Volts
XX.X
< 6.0 FAIL (red)
<10 T Chg (yellow)
10-15.5 (Green)
Vdc
10.0 to 15.5
Battery 2 Volts
XX.X
< 6.0 FAIL (red)
<10 T Chg (yellow)
10-15.5 (Green)
Vdc
10.0 to 15.5
Calc Battery Time
XX
Min
0 to 30
Date Battery Tested
XX-ABC-XXXX
Last Full Discharge Time
XXX
Min
Ext’l Battery Current
X.XX
A
Summary of Contents for Engstrom Carestation
Page 1: ...Engström Carestation Technical Reference Manual ...
Page 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Page 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Page 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Page 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Page 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...