7 Troubleshooting
1505-1018-000 08/07
7-25
Neo flow sensor
off
Neo Flow Sensor Off
Medium Neo flow sensor turned off in neo flow
sensor setup menu.
• Turn on neo flow sensor.
Remove airway
module
Neo module not
compatible
Low
Any MGAS module is installed in the
module bay during neo patient type.
• Remove module or change patient type to adult or pediatric.
No neo flow
sensor
No Neonatal Flow
Sensor - Standby
Low
Both neonatal flow sensor hot wires
reporting less than 20 ADC as
measured by the VCB or VMB.
• Calibrate neo flow sensor.
• Check neo flow sensor and cable connections.
• Replace neo flow sensor.
• Replace neo flow sensor cable.
• Replace VMB.
• Repace motherboard.
• Replace VCB.
Neo flow sensor
reversed
Neo Flow Sensor
Reversed
High
Neo TVinsp more negative than -3 mL
for 6 consecutive mechanical breaths.
• Verify FMI 34014 has been
completed on unit.
• Calibrate neo flow sensor.
• Check neo flow sensor and cable connections.
• Replace neo flow sensor.
• Replace neo flow sensor cable.
• Replace VMB.
• Repace motherboard.
• Replace VCB.
No neo flow
sensor
No Neonatal Flow
Sensor - Therapy
High
Both neonatal flow sensor hot wires
reporting less than 20 ADC as
measured by the VCB or VMB.
• Calibrate neo flow sensor.
• Check neo flow sensor and cable connections.
• Replace neo flow sensor. Neonatal flow sensor is warranteed for 90 days.
• Replace neo flow sensor cable.
• Replace VMB.
• Repace motherboard.
• Replace VCB.
No battery
backup
Degraded Battery
Medium
The total battery voltage is <20VDC
-or-
Each battery voltage is <10VDC
• Verify unit has been connected to mains for more than 24 hours.
• Check error log for related messages.
• Use Service App to determine charge state. If in trickle charge, replace batteries.
• Check/replace harness.
• Check/replace PMB.
Alarm message
Alarm ID
Priority
Alarm Condition
Special Behavior/Comments
• Action/Troubleshooting
Summary of Contents for Engstrom Carestation
Page 1: ...Engström Carestation Technical Reference Manual ...
Page 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Page 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Page 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Page 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Page 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...