Engström Carestation
8-24
08/07 1505-1018-000
8.7.2 VMB outputs
The VMB Outputs group controls items that can be turned on or off. Items
whose default state is “On”, initially have a checkmark in the checkbox. For
items whose default state is “Off”, the checkbox is initially unmarked.
8.7.3 VMB data
The VMB Data group shows the reported values for Expiratory Flow and
O
2
concentration.
Item
Function
Default
Exh Valve Pwr
Opens/closes the exhalation valve (opens when checked). Not
checked
Safe Valve
Energized
Opens/closes the safety valve from the VMB
(energized when checked).
Not
checked
O2 Sensor Off
Turn oxygen sensor On (not checked) or Off (checked)
Not
checked
Exh Flow Sensor
Pwr
Turn exhalation sensor On (checked) or Off (not checked)
Checked
VMB Buzzer On
The VMB turns the buzzer On/Off.
Not
checked
VMB Reset
Resets the VMB (disables the watchdog) when checked.
Not
checked
Exp Press Zero
Open
Opens/Closes the expiratory pressure zero valve. The
valve will be opened when checked.
Not
checked
Item
Description
Format
Units
Exp Flow
Expiratory flow sensor measurement.
XXX.XX
mL/min
O2 %
Measured oxygen concentration from the
oxygen sensor.
XXX
%
Summary of Contents for Engstrom Carestation
Page 1: ...Engström Carestation Technical Reference Manual ...
Page 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Page 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Page 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Page 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Page 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...