Engström Carestation
5-8
08/07 1505-1018-000
5.2.5 Vent engine leak
test (low pressure)
There are two low pressure leak tests to perform: P1 and P2.
Low P1 Test
1. Verify that both shut-off valves are closed.
2. On the VCB screen:
• Energize (
!
) the
Aux
iliary
Press
ure
Purge
valve.
• Verify that the
Air Flow Valve
,
O2 Flow Valve
and the
Exhal
ation
Flow
Valve
are energized (
!
).
• Set the
Air DAC
,
O2 DAC
, and the
Exh
alation
DAC
counts to 60,000
(this will release all pressurized gas from the system).
3. Attach the “low-pressure test tool” to the
INSP
iratory outlet.
4. Plug the following ports:
• Expiratory inlet (
EXP
).
• Auxiliary pressure port (
Paux
).
5. Using the low pressure test tool, increase the pressure in the low pressure
side of the system until the measured
Expiratory
Press
ure (VMB),
Auxiliary
Press
ure (VCB), and
Inspiratory
Press
ure (VCB) readings indicate 70
cmH2O.
Note: Ensure that the three pressures are stable and approximately equal
(within 5 cmH2O). If not, refer to Section 7.2.2.
6. Once 70 cmH2O is reached, clamp the tube on the low pressure test tool
to prevent gas flow from the
INSP
iratory outlet.
7. On the VCB screen, select
Low P Leak
1
1
1
1
to start the low pressure P1leak
test.
8. After the test is complete (15 seconds):
• If the leak result does not exceed 10 ml/min, proceed to to step 9.
(Note: The Low P2 test is not required if Leak result ≤ 10 ml/min.)
• If Leak result is >10 ml/min and ≤ 35 ml/min, perform the Low P2 test.
• If the Leak rate exceeds 35 ml/min, refer to Section 7.2 to help identify
the cause of the leak.
9. Remove the plugs from the
EXP
iratory inlet and the
Paux
port.
10.De-energize (
"
) the
Aux
iliary
Press
ure
Purge
valve.
AB.98.088
Summary of Contents for Engstrom Carestation
Page 1: ...Engström Carestation Technical Reference Manual ...
Page 12: ...Engström Carestation x 08 07 1505 1018 000 ...
Page 52: ...Engström Carestation 2 26 08 07 1505 1018 000 ...
Page 100: ...Engström Carestation 5 26 08 07 1505 1018 000 ...
Page 124: ...Engström Carestation 7 16 08 07 1505 1018 000 ...
Page 188: ...Engström Carestation 8 26 08 07 1505 1018 000 ...