background image

Ultrasonic Nebulizer

NE-U780

(NE-U780-E)

Instruction Manual

IM-NE-U780-E-01-11/2015

5371508-1B

 This product is a medical device and should be used under the guidance of a physician. For type, dose, 

and regime of medication, follow the instructions of your doctor or pharmacist.

 This product is not designed to function as a home-use inhaler for the purpose of moistening the mucous 

membranes of the nose and throat.

EN

FR

DE

IT

ES

NL

RU

AR

Table of Contents

Important Information

Intended Use ...............................1
Important Safety Precautions ......2
Exemptions ..................................3

Important 

Information

Know Your Unit

Product Features .........................4
Packaging Contents ....................5
Display .........................................5
Part Names .................................6

Know Y

our 

Unit

Operating Instructions

Before Inhalation .........................7
Inhalation Method ......................10
Inhalation Method  
(using optional accessories) ......13

Operating 

Instructions

Care and Maintenance

Maintenance ..............................16
Disinfecting ................................19
Storage ......................................21

Care and 

Maintenance

Technical Data, Warranty, etc.

Troubleshooting .........................22
Technical Data ...........................23
Optional Medical Accessories ...25
Warranty ....................................26

Te

chnical Data, 

W

arranty

, etc.

Summary of Contents for NE-U780

Page 1: ...throat EN FR DE IT ES NL RU AR Table of Contents Important Information Intended Use 1 Important Safety Precautions 2 Exemptions 3 Important Information Know Your Unit Product Features 4 Packaging Cont...

Page 2: ...of a medical expert Some hospitals clinics may rent out the devices Durable period Durable periods are as follows provided the product is used to nebulize 30 ml of saline solution at 26 C with 26 C w...

Page 3: ...ain unit contact are not contaminated Make sure that the air filter is correctly attached To avoid the medication residue on the face be sure to wipe the face after removing the mask Do not get nebuli...

Page 4: ...failure or damage or accidents involving the product resulting from a failure to adhere to the safety precautions and operating instructions described in this manual 5 Product failure or damage result...

Page 5: ...E ZKLFK WKH medication becomes an aerosol 1HEXOL DWLRQ 7KH SURFHVV E ZKLFK WKH DWRPL HG medication is expelled by the product 1 Smooth shape for easy cleaning The main unit is easy to clean and disin...

Page 6: ...hen using the timer Indicates the remaining nebulization time The time can be set from 1 to 30 min During continuous nebulization The digital portion of the display cycles through a simple animation P...

Page 7: ...f main unit Top of main unit Fuse Electric power socket Power cord Display Handle Mouthpiece Inhalation hose Fan cover Fan Air filter case Air filter Vibrator Nebulization set Medication cup cover Sil...

Page 8: ...acing your finger under the protrusion on the medication cup holder remove the nebulization set Protrusion 3 Fill the removable water tank with tap water Fill to the optimal water level taking care no...

Page 9: ...to the medication cup Min 10 ml Max 150 ml Notes HIRUH DGGLQJ PHGLFDWLRQ PDNH VXUH WKH medication cup is clean not damaged or misshapen I XVLQJ D V ULQJH WDNH FDUH QRW WR GDPDJH WKH medication cup wit...

Page 10: ...power cord 4 Turn on the power button and make sure all the Error LEDs on the error display light up and then turn back off If one or more of the Error LEDs remain lit see Troubleshooting Page 22 for...

Page 11: ...ULIZATION VOLUME adjustment buttons set the air flow volume and nebulization volume You can check both settings with the level displays which change to reflect the current value of each The air flow v...

Page 12: ...ion mask optional can be used instead of the mouthpiece During inhalation Do not bend the inhalation hose or block the mouthpiece It may cause medication to accumulate in the hose reducing the nebuliz...

Page 13: ...mple display with the timer set to 10 min When operation stops After about 5 sec This prepares the time for the next patient To stop inhalation while using the timer Press the START STOP button to sto...

Page 14: ...fan cover for bacterial filter Instead of the normal fan cover use the optional NEB BFC 78E Fan Cover for Bacterial Filter 2 Remove the air volume adjustment lever on the long time nebulization set an...

Page 15: ...an be connected to the product Continuous nebulization of medication volumes in excess of 150 ml can be accomplished by using the optional NEB LMCC 78E Long time Nebulization set and U10 11 E Supply M...

Page 16: ...is not bent Medication will not be supplied if the tube is bent If the tube is too long cut it to a suitable length For subsequent steps see No 2 under Step 3 in the Before Inhalation section Page 9 S...

Page 17: ...mouthpiece and other accessories from the main unit 3 Remove the removable water tank from the main unit 4 Remove the nebulization set from the removable water tank and disassemble it 5 Dispose of an...

Page 18: ...e air filter is not contaminated If the air filter is excessively contaminated replace it with a new air filter 4 5 Return the air filter case and fan cover to their original positions and align the m...

Page 19: ...t clean 5 Reattach the fan air filter case and fan cover When reattaching the fan push it firmly all the way onto the fan installation axis 6 Align the medication cup cover so that it is once more con...

Page 20: ...infectant see its instruction manual and precautionary information Some disinfectants may trigger an allergic reaction Although the amount of time for which parts should be immersed varies with the di...

Page 21: ...diene styrene resin ABS Poly phenylene sulfide resin PPS Polyethylene resin PE Polycarbonate resin PC Polymethylpentene PMP Polypropylene resin PP Polystyrene resin PS Polyvinyl chloride resin PVC The...

Page 22: ...icon rubber Silicon packing Washer Silicon rubber Bottle PP Silicon packing Washer Silicon rubber Tube Silicon rubber Hanger Stainless steel Long time Nebulization set Supply Medication set Do not dis...

Page 23: ...the nebulization volume or air flow volume setting too low Increase the setting Page 10 Is the room temperature or water temperature too low Replace the water in the removable water tank with water w...

Page 24: ...tank Medication cup cover Medication cup holder Inhalation hose M Mouthpiece Medication cup 2 pcs Power cord Instruction Manual including product warranty Classifications Class I Protection against e...

Page 25: ...th the aim to prevent unsafe product situations the EN60601 1 2 2007 standard has been implemented This standard defines the levels of immunity to electromagnetic interferences as well as maximum leve...

Page 26: ...aminants out of the airflow Bacterial Filter Materials PP PMP Model U17 2 E Remarks Includes adapter Used for inhalation when cleansing the respiratory tract etc Cannot be reused Fan cover for bacteri...

Page 27: ...rt b Costs for repairs and or defects resulting from repairs done by unauthorised persons c Periodic check ups and maintenance d Failure or wear of accessories or other attachments other than the main...

Page 28: ...EALTHCARE Co Ltd Matsusaka Factory 1855 370 Kubo cho Matsusaka shi Mie 515 8503 Japan Subsidiaries OMRON HEALTHCARE UK LTD Opal Drive Fox Milne Milton Keynes MK15 0DG UK www omron healthcare com OMRON...

Reviews: