20
EN
Important
Information
Know Y
our
Unit
Operating
Instructions
Care and
Maintenance
Te
chnical Data,
W
arranty
, etc.
Disinfecting
●
Disinfection chart
Parts can be disinfected using the methods listed in the chart below. For more information about
disinfecting the main unit, see “Disinfecting the main unit”. (
Page 19) For more information
about the disinfection process, see the disinfectant’s instructions and precautions.
Note: Materials are indicated by the following abbreviations:
Acrylonitrile butadiene styrene resin: ABS, Poly phenylene sulfide resin: PPS, Polyethylene resin: PE,
Polycarbonate resin: PC, Polymethylpentene: PMP, Polypropylene resin: PP, Polystyrene resin: PS,
Polyvinyl chloride resin: PVC, Thermoplastic elastomer: TPE
Medication cup cover
(with silicon packing
(Washer))
Medication cup
Removable water tank
Note: Disassemble the
handle
cover
before
disinfection.
Fan cover
Medication cup holder
PP
PP
ABS
Parts
Materials
Chemical
disinfectant
Boiling
water
Autoclave
Fan cover for
bacterial filter
ABS
PP
PS
Air filter
Fan
Air filter case
Mouthpiece
PP
Medication cup cover: PMP
Silicon packing (Washer):
Silicon
rubber
Removable water tank:
ABS/PPS
Handle cover: ABS
PP
: applicable : not applicable