Troubleshooting
Issue
Inspection location
Solution
The power LED
does not light up.
Is the power cord connected correctly to a power
outlet and the main unit?
Plug the product into an outlet.
(
Page 9)
Nebulization is not
functioning.
Is the “
WATER LEVEL” Error LED lit up due to
there being too little water in the removable water tank?
Fill the tank with water to the optimal water
level.
(
Page 7)
Have the medication cup and removable water tank
been attached properly?
Attach the parts properly.
(
Page 8)
Is the fan cover open?
Close the cover tightly.
Has the fan been attached?
Attach the fan properly.
(
Page 18)
Is there too much medication in the medication cup?
Reduce the amount of medication to 150 ml or
less.
(
Page 8)
Did you add distilled or purified water to the
removable water tank?
Add tap water to the tank.
Are the contacts on the main unit or removable water
tank contaminated?
Wipe away any contamination with a cloth that
has been moistened with water or sanitizing
alcohol and then wrung out lightly.
Is the inhalation hose bent, causing medication to
accumulate inside it?
Straighten the inhalation hose and remove any
accumulated medication.
Has the START/STOP button been turned on?
Press the START/STOP button. ( Page 11)
The aerosol
volume is low.
The nebulization
capacity is low.
Is there too little medication in the medication cup?
Add more medication so that there is at least
10 ml in the cup.
(
Page 8)
Is the medication cup damaged or misshapen?
Replace the medication cup with a new one.
Is there too much medication in the medication cup?
Reduce the amount of medication so that it is
150 ml or less.
(
Page 8)
Is the nebulization volume or air-flow volume setting
too low?
Increase the setting.
(
Page 10)
Is the room temperature or water temperature too
low?
Replace the water in the removable water tank
with water whose temperature is about 26 °C.
Is the medication cup contaminated?
Disinfect the medication cup. If unable to
remove the contamination, replace the
medication cup with a new one.
(
Page 16 and Page 20)
Nebulization is
inconsistent.
Is there too much medication in the medication cup?
Reduce the amount of medication so that it is
150 ml or less.
(
Page 8)
Does the medication cup contain 150 ml of
medication, and are the maximum values for both the
air-flow volume and nebulization volume settings
being used?
Adjust the nebulization volume. ( Page 10)
Issue
Inspection location
Solution
Error LEDs
LED
is lit up.
There is too little water in the
removable water tank.
Fill the tank with water to the optimal
water level.
(
Page 7)
LED
is lit up.
The medication cup cover is open.
Close the cover tightly. ( Page 8)
LED
is lit up.
The fan cover is open.
Close the cover tightly.
LED
is flashing.
It’s time to inspect the air filter.
(The LED lights up after about 100
hours of nebulization.)
Inspect the air filter. If it is excessively
contaminated, replace it with a new air filter.
(
Page 17)
E1 indicator
The main unit’s internal fan has
stopped.
Unplug the cord, label the main unit to
indicate that it is broken, and contact an
authorised OMRON retail outlet or
distributor.
E3 indicator
The main unit’s internal software
experienced an error.
E4 indicator
There is a short in the main unit’s
internal oscillation circuit.
E5 indicator
The removable water tank has not
been secured properly.
Remove the removable water tank and
then secure it in the main unit.
(
Page 9)
Note: If the suggested remedy does not solve the problem, do not try to repair the device - the device is not user serviceable.
Return the device to an authorised OMRON retail outlet or distributor.