PQRM5100 11 Ux Ix xx xx (PS)
74
20200205-V1
Analogue outputs (optional)
Output type:
2 galvanic isolated active current outputs
(configurable, scalable)
Ranges:
0-20 mA / 4-20 mA or
0-5 mA / 1-5 mA
Burden: 500
ohm
(max.)
Refreshing time:
same as the measuring time (100 ms)
Setting time: (10-90%)
Max. 60 ms
Overcurrent: 20.8
mA
Error:
<4 uA (23 °C ±2 °C), <40 uA (-20 - 60 °C)
Burden resistance effect:
practically zero
Pulse outputs (optional):
Output type:
2 galvanic isolated transistor
Rating:
30 V, 50 mA
Power supply:
Supply voltage:
24 VDC ±10%
PQRM5100 11 Ux Ix xx xx
or
230 V AC/DC ±10%
PQRM5100 11 Ux Ix xx xx PS
Power consumption:
1.5 VA / 1 W
Galvanic isolation:
Operating isolation voltage:
250 Veff (between measuring inputs and
power supply input)
Test /Type voltage:
4200 VDC (1 min.) (between measuring
inputs and power supply input)
500 VDC (between output-power supply
terminals)
Capacity:
20 pF (between input, output, power supply
terminals)
Protection class:
II. reinforced insulation
Leakage current:
0,005 mA (between input, output, power
supply terminals)
Ambient conditions:
Operating temperature range:
0-60 °C
Storage temperature range:
0-70 °C
Relative humidity:
90 % (max., non condensing)
Place of installation:
cabinet
Summary of Contents for PQRM5100 11 U I Series
Page 1: ...PQRM5100 11 Ux Ix xx xx PS Single Phase power transmitter Instruction manual...
Page 76: ...PQRM5100 11 Ux Ix xx xx PS 76 20200205 V1 9 2 Application examples...
Page 77: ...PQRM5100 11 Ux Ix xx xx PS 20200205 V1 77...
Page 78: ...PQRM5100 11 Ux Ix xx xx PS 78 20200205 V1 RS485 bus topology...
Page 79: ...PQRM5100 11 Ux Ix xx xx PS 20200205 V1 79...
Page 80: ......