Spyder3 Color Manual
15
DALSA
03-032-20059-01
Table 9: DALSA Camera Control Configuration
Signal
Configuration
CC1
EXSYNC
CC2
PRIN
CC3
Direction
See Appendix B for the complete DALSA Camera Link configuration table, and refer to
the DALSA Web site for the official Camera Link documents
(
http://www.dalsa.com/mv/knowledge/appnotes.aspx
).
Input Signals, Camera Link
The camera accepts control inputs through the Camera Link MDR26F connector.
The camera ships in internal sync, maximum exposure time (exposure mode 7).
EXSYNC (Triggers Line Readout)
Line rate can be set internally using the serial interface. The external control signal
EXSYNC is optional and enabled through the serial interface. This camera uses the
falling
edge of EXSYNC
to trigger pixel readout. See Setting the Exposure Mode, page 32, for
details on how to set line rates, exposure times, and camera modes.
Output Signals, Camera Link
These signals indicate when data is valid, allowing you to clock the data from the camera
to your acquisition system. These signals are part of the Camera Link configuration and
you should refer to the DALSA Camera Link Implementation Road Map for the standard
location of these signals (
http://www.dalsa.com/mv/knowledge/appnotes.aspx
).
Clocking Signal
Indicates
LVAL (high)
Outputting valid line
DVAL (high)
Valid data (unused, tied high)
STROBE (rising edge)
Valid data
FVAL (high)
Outputting valid frame (unused, tied high)
The camera internally digitizes 12 bits and outputs the 8 MSB or all 12 bits depending on
the camera‘s Camera Link operating mode.
i
Summary of Contents for Spyder3 SC-30-02K80-00-R
Page 24: ...24 Spyder3 Color Manual 03 032 20059 01 DALSA ...
Page 78: ...78 Spyder3 Color Manual 03 032 20059 01 DALSA ...
Page 90: ...90 Spyder3 Color Manual 03 032 20059 01 DALSA ...
Page 98: ...98 Spyder3 Color Manual 03 032 20059 01 DALSA ...
Page 104: ...104 Spyder3 Color Manual 03 032 20059 01 DALSA ...
Page 108: ...108 Spyder3 Color Manual 03 032 20059 01 DALSA ...