Electric power
Property
Value
Supply voltage
100-120/220-240 V AC ± 10%, 50 to 60 Hz
Phases
Single
Max power
300 VA
Ingress protection
IP 21
Fuses
2x T 3.15AL 250 V
Over voltage category
II
Polution degree
2
Operating limits
Parts and properties
Value
Feed pump
Pressure at 20°C
Max 4.0 bar pressure alarm/cut-out
Transfer pump
Feed temperature
2°C to 50°C
Outlet pressure
Max 1 bar
Permeate pump
Permeate tempera-
ture
2°C to 50°C
Inlet pressure
Max 1 bar
Outlet pressure
Max 1 bar
Environmental ranges
Ambient temperature
2°C to 35°C
Temperature, trans-
port
-25°C to +60°C
Temperature gradients of ≤ ± 1°C/min
Dry instrument (drained fluid path)
Temperature, storage
-25°C to +50°C
Temperature gradients of ≤ ± 1°C/min
Dry instrument (drained fluid path)
Air humidity
Max. relative humidity 80% for tempera-
tures up to 31°C, decreasing linearity to
50% relative humidity at 40°C
Altitude
Max 2000 m
Liquid temperature
Process temperature
2°C to 40°C
11 Reference information
11.1 Specifications
184
ÄKTA flux s Operating Instructions 29046893 AG