Step
Action
5
Note the
Pp permeate pressure sensor value shown on the indicator panel.
Collect fluid between time point 0 and time point T. Measure the mass of the
fluid in beaker B with an external balance and calculate the flow rate (weight/
time).
6
Perform the corresponding procedure (step 2 to 6) at a high RPM value of the
permeate pump.
7
Tap
Settings on the main screen.
8
Tap
Calibration on the screen that appears.
9
Tap
Permeate pump on the Calibration Settings screen.
6 Run preparations
6.4 Calibration
6.4.3 Calibration of the permeate pump
122
ÄKTA flux s Operating Instructions 29046893 AG