6.4.3
Calibration of the permeate pump
Follow the instructions below to calibrate the system flow for the permeate pump. The
permeate pump is optional.
Step
Action
1
Connect tubing to the permeate pressure sensor inlet (1) and the permeate
pump outlet (2).
2
Fill beaker A with process liquid and put the tubing in the beaker.
3
Set a low RPM value of the permeate pump on the control panel and tap
Start. Let the pump run for a couple of minutes.
For an instruction how to use set-points, see
Adjust set-points, on page 161
.
6 Run preparations
6.4 Calibration
6.4.3 Calibration of the permeate pump
120
ÄKTA flux s Operating Instructions 29046893 AG