6.4
Calibration
Introduction
This section describes the procedures used to perform calibration of the pumps and
other parts in ÄKTA flux s.
When to use these procedures
It is recommended to perform calibrations before each run.
Allowed user performed calibrations:
•
Pump flow rate for installed pumps
•
Tank level
•
Re-zero function of installed pressure sensors
•
Motor RPM range calibration for installed motors (pumps and mixer)
Allowed service performed calibrations:
•
All user allowed calibration
•
Full range calibration of installed pressure sensors
The pressure and tubing dimension used when calibrating the system should always
correspond to the pressure and tubing dimension used when performing a run.
In this section
Section
See page
Calibration of the transfer pump
Calibration of the permeate pump
Calibration of the pressure sensors
6 Run preparations
6.4 Calibration
ÄKTA flux s Operating Instructions 29046893 AG
109