6.4.5
Motor control
A motor control calibration is performed in order to recognize which motors that are
connected to the system and shall always be performed when the optional transfer or
permeate pump is installed. Follow the instructions below to perform the motor
control.
Step
Action
1
Tap
Settings on the main screen.
2
Tap
Calibration to open Calibration Settings screen.
3
Tap
Motor Control (1).
1
2
4
Tap (2) the
Feed Pump, Permeate Pump, Transfer pump, or the Mixer,
depending on which of the motors you intend to calibrate.
Result:
The selected motor calibrates itself by starting on the lowest rotation speed
and accelerating to the highest rotation speed.
6 Run preparations
6.4 Calibration
6.4.5 Motor control
126
ÄKTA flux s Operating Instructions 29046893 AG