6.4.6
Tank level
Tank balance
It is recommended to calibrate the tank balance prior to each run to obtain a reliable
output. The tank balance must also be calibrated every time the system has been
moved, or if there has been any interference with the balance.
Note:
Do not interfere with the tank balance during the calibration procedure and
filter run.
Calibrate the tank level
Step
Action
1
Tap
Settings on the main screen.
2
Tap
Calibration to open Calibration Settings.
3
Tap
Tank Level.
4
Empty the tank.
6 Run preparations
6.4 Calibration
6.4.6 Tank level
ÄKTA flux s Operating Instructions 29046893 AG
127